ZRA function in CS potentiostat/Galvanostat

Source:Corrtest Time:2021-12-02 View:0

 CS series Potentiostat/Galvanostat is all equipped with the function of ZRA(zero resistance ammeter) where no voltage or current is needed to apply on the measuring system.
Therefore, ZRA can also be called Open Circuit Technique.




The electrode connection way in ZRA is different from that of general electrochemical testing. In ZRA the connection way is as follows: 
- WE (green) clamp connects to the sample.

 - The GND lead(black) clamp connects to another sample.
 - RE(yellow) clamp connects to reference electrode.
 - CE(red) clamp is  not used.

 ZRA is usually applied in Galvanic corrosion and electrochemical noise (EN). 