ENGLISH
中文
Email: sales2@corrtest.com.cn
Phone:+86 13469965984

FAQs

How should I know if the reference electrode i'm using is good or need replacement?

Source:Corrtest Time:2021-12-02 View:0
Prepare a new reference electrode (same type as the reference electrode you’re using, eg both Ag/AgCl RE).
Then put the two electrodes(this new and the one you're using) in a small beaker containing saturated KCl solution, and measure the potential difference between them by a potentiostat or digital voltmeter. If the potential difference is higher than 10mV, you should refresh the filling solution of the electrode you’re using at once, or directly change for a new one.


 

About Us   Potentiostat/Galvanostat   Accessories   Support   Contact Us  

Copyright By © 2008-2024 Wuhan Corrtest Instrument Corp., Ltd

This website stores cookies on your computer. These cookies are used to improve your website experience and provide more personalized services to you, both on this website and through other media. To find out more about the cookies we use, see our Privacy Policy.
We won't track your information when you visit our site. But in order to comply with your preferences, we'll have to use just one tiny cookie so that you're not asked to make this choice again.